Configuring DICOM Query/Retrieve
================================

Starting with release 0.7.0, Orthanc supports DICOM Query/Retrieve
(i.e. Orthanc can act as C-Find SCP and C-Move SCP). To configure this
feature, follow these instructions:

* Get the AET (Application Entity Title), the IP address and the port
  number of your DICOM client.
* Add an entry in the ``DicomModalities`` section of your Orthanc
  :ref:`configuration file <configuration>` to reflect these settings.
* Restart Orthanc with the updated configuration file.
* Symmetrically, in your DICOM client, configure an additional DICOM
  node corresponding to your Orthanc AET (see the ``DicomAet``
  parameter of your Orthanc configuration, by default, ORTHANC), IP
  address et port number (cf. ``DicomPort``, by default 4242).

If you encounter configuration problem with query/retrieve, please be
sure to read the :ref:`dicom-move` section.


.. _ginkgo-cadx:

Illustration: Ginkgo CADx
-------------------------

This section explains how to connect the `Ginkgo CADx viewer
<http://ginkgo-cadx.com/en/>`__ to an Orthanc server, both running on
the local computer. First, start Orthanc without specifying a
configuration file (i.e. using its default configuration).  Then, open
the settings pannel of Ginkgo CADx:

.. image:: ../images/GinkgoCADx1.png
           :align: center
           :width: 500px

Write down the AET and the DICOM port of Ginkgo CADx (in this case,
respectively ``GINKGO_001`` and ``11112``). Then, create a new DICOM
node corresponding to Orthanc by clicking on the "*New...*" button,
which raises the following dialog box:

.. image:: ../images/GinkgoCADx2.png
           :align: center
           :width: 400px

.. highlight:: json

Complete the highlighted fields with the default parameters of
Orthanc, as shown in the screenshot.  Do not forget to click on the
"*Test connection*" button to make sure the :ref:`C-Echo <dicom-echo>`
succeeds. Stop Orthanc, create a new :ref:`configuration file
<configuration>`, and declare Gingko CADx in the ``DicomModalities``
section, given the parameters you wrote down above::

  // The list of the known DICOM modalities
  "DicomModalities" : {
    "ginkgo" : [ "GINKGO_001", "localhost", 11112 ]
  },

Restart Orthanc using this newly created configuration file, and
you will be able to do query/retrieve from Gingko CADx through the
"*PACS Q/R*" menu.