Mercurial > hg > orthanc-book
diff Sphinx/source/faq/query-retrieve.rst @ 0:901e8961f46e
initial commit
author | Sebastien Jodogne <s.jodogne@gmail.com> |
---|---|
date | Fri, 22 Apr 2016 12:57:38 +0200 |
parents | |
children | 8f3f79aeb8cc |
line wrap: on
line diff
--- /dev/null Thu Jan 01 00:00:00 1970 +0000 +++ b/Sphinx/source/faq/query-retrieve.rst Fri Apr 22 12:57:38 2016 +0200 @@ -0,0 +1,64 @@ +.. _query-retrieve: + +Configuring DICOM Query/Retrieve +================================ + +Starting with release 0.7.0, Orthanc supports DICOM Query/Retrieve +(i.e. Orthanc can act as C-Find SCP and C-Move SCP). To configure this +feature, follow these instructions: + +* Get the AET (Application Entity Title), the IP address and the port + number of your DICOM client. +* Add an entry in the ``DicomModalities`` section of your Orthanc + :ref:`configuration file <configuration>` to reflect these settings. +* Restart Orthanc with the updated configuration file. +* Symmetrically, in your DICOM client, configure an additional DICOM + node corresponding to your Orthanc AET (see the ``DicomAet`` + parameter of your Orthanc configuration, by default, ORTHANC), IP + address et port number (cf. ``DicomPort``, by default 4242). + +If you encounter configuration problem with query/retrieve, please be +sure to read the :ref:`dicom-move` section. + + +.. _ginkgo-cadx: + +Illustration: Ginkgo CADx +------------------------- + +This section explains how to connect the `Ginkgo CADx viewer +<http://ginkgo-cadx.com/en/>`__ to an Orthanc server, both running on +the local computer. First, start Orthanc without specifying a +configuration file (i.e. using its default configuration). Then, open +the settings pannel of Ginkgo CADx: + +.. image:: ../images/GinkgoCADx1.png + :align: center + :width: 500px + +Write down the AET and the DICOM port of Ginkgo CADx (in this case, +respectively ``GINKGO_001`` and ``11112``). Then, create a new DICOM +node corresponding to Orthanc by clicking on the "*New...*" button, +which raises the following dialog box: + +.. image:: ../images/GinkgoCADx2.png + :align: center + :width: 400px + +.. highlight:: json + +Complete the highlighted fields with the default parameters of +Orthanc, as shown in the screenshot. Do not forget to click on the +"*Test connection*" button to make sure the :ref:`C-Echo <dicom-echo>` +succeeds. Stop Orthanc, create a new :ref:`configuration file +<configuration>`, and declare Gingko CADx in the ``DicomModalities`` +section, given the parameters you wrote down above:: + + // The list of the known DICOM modalities + "DicomModalities" : { + "ginkgo" : [ "GINKGO_001", "localhost", 11112 ] + }, + +Restart Orthanc using this newly created configuration file, and +you will be able to do query/retrieve from Gingko CADx through the +"*PACS Q/R*" menu.