Message300

Author admin
Recipients
Date 2017-09-12.00:25:01
Content
[BitBucket user: Денис Смирнов]
[BitBucket date: 2017-09-11.22:25:01]

When I make a simple C-Move request to Orthanc from a workstation Orthanc establishes back an association and makes C-Store responce with a zero Move Originator Message ID (0000, 1031) value. According to DICOM NEMA Move Originator Message ID contains the Message ID (0000,0110) of the C-MOVE-RQ Message from which this C-STORE sub-operation is being performed. But Orthanc always returns zero and it causes errors when user tries to view images directly from Orthanc in Philips Portal. Here is an example DICOM output
![Снимок экрана_2017-09-12_08-23-09.png](1076252711-%D0%A1%D0%BD%D0%B8%D0%BC%D0%BE%D0%BA%20%D1%8D%D0%BA%D1%80%D0%B0%D0%BD%D0%B0_2017-09-12_08-23-09.png)

I was given an advice in [Orthanc users google group](https://groups.google.com/forum/#!topic/orthanc-users/oHoKGB2Az0I) to open an issue about this problem.